Shanghai Haozhuang Instrument Co., Ltd. has developed the Haozhuang (LNB) brand shock light incubator, which is energy-saving and environmentally friendly, leading the country. Professional manufacturing is more secure and affordable
Analysis of common failures of light incubator
1. The compressor does not refrigerate; (1) The compressor is always in working state under the balance mode, the cooling effect is better and the compressor can be effectively maintained. ⑵, if it is a single-system operation, it will work only when the temperature drops to a certain value, and it will cause a large temperature fluctuation when the door is opened and closed. The compressor is started frequently, and it is easy to shorten the service life of the compressor.
2. Uneven temperature in the incubator; etc. (1) It may be caused by excessively dense discharge items or the circulation fan in the box will not work;
3. Circulation tuyere freezes. Each refrigeration instrument will have a defrost cycle, during which the compressor will not refrigerate. The heating element melts the ice on the condenser. The melted water is likely to gather in the air outlet and eventually cause ice.
Maintenance and cultivation of light incubator:
1. The shell of the incubator should be reliably grounded.
2. In order to ensure the effective heat dissipation of the condenser, the distance between the condenser and the wall should be greater than 100mm. There should be a gap of 50mm on the side of the box and at least 300mm of space on the top of the box.
3. The incubator should be placed in a cool, dry, well-ventilated place away from heat sources and sunlight. Place it steadily to prevent vibration and noise.
4. When carrying, repairing and maintaining the incubator, avoid collision, shaking and vibration; the maximum inclination is less than 45 degrees.
5. The instrument does not work suddenly, please check whether the fuse tube (behind the box) is burnt out, and check the power supply.
6. When cooling the incubator, it is not advisable to make the difference between the temperature inside the box and the ambient temperature greater than 25 degrees.
Shanghai Haozhuang Instrument Co., Ltd. is a laboratory instrument and equipment company integrating R & D, production and sales. Develop Hao Zhuang (LNB) brand instruments, the main products are low temperature thermostat, coolant circulation pump, constant temperature water tank, high temperature circulator, constant temperature shaker, freeze dryer, snowflake ice machine, ultrasonic cleaning machine, ultrasonic cell crusher, Ultrasonic extraction machine, sterile homogenizer, ultra-clean workbench and other instruments.